Author Affiliations
Abstract
We develop the kinematic model of portable coordinate measuring arm (PCMA) based on the research on its mechanical characteristics and working principle, and present a comprehensive thermally-induced error correction model for PCMA. Based on the model and nominal data provided by a highly accurate coordinate measuring machine, the technique for thermally-induced error compensation of PCMA is presented along with measuring a standard artifact at the corresponding temperature field. Experimental results demonstrate that the deviation of PCMA at different temperatures is controlled in the range of ±10 μm using this technique to calibrate the PCMA.
120.3940 Metrology 120.3930 Metrological instrumentation 
Chinese Optics Letters
2014, 12(s1): S11203
Author Affiliations
Abstract
1 State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072
2 Precision Engineering Institute, He'nan Polytechnic University, Jiaozuo 454003
A technique to calibrate faro retro probe using laser tracker system (LTS) is presented. The deep-access retro probe enables LTS to perform three-dimensional (3D) measurements of surface hidden features. It has been shown that misalignment errors are the key contributor to measuring errors of retro probe. A device for measuring misalignment errors of retro probe is invented. Theoretical analysis and experiment show that using this technique to calibrate retro probe, the misalignment errors of retro probe can be eliminated effectively and quickly.
激光跟踪系统 隐藏特征测量探头 不对称误差 校准 120.0120 Instrumentation, measurement, and metrology 140.0140 Lasers and laser optics 230.0230 Optical devices 
Chinese Optics Letters
2007, 5(5): 281

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!